HomeAbout UsProductsContact Us
 


Temperature Surface Calibrator


Anritsu Model ACSII series products are temperature calibration systems for surface temperature and internal temperature measuring devices. Temperatures can be set across a wide range, from room temperature to 500¢XC. These calibration systems are designed for calibration or to check temperature measuring probes.

Technical Specs

Model ACS II-1000 ACS II-2000 ACS II-3000 ACS II-4000
Calibration Object Surface Internal Surface Internal
Range Rt to 100°C +0.5 / -0.7°C +0.5 / -0.7°C ±0.26°C +0.4 / -0.6°C ±0.26°C
100 to 200°C +0.5 / -1.1°C +0.5 / -1.1°C ±0.32°C +0.3 / -0.9°C ±0.32°C
200 to 300°C +0.5 / -1.5°C +0.5 / -1.5°C ±0.37°C +0.1 / -1.1°C ±0.37°C
300 to 400°C - +0.4 / -2.0°C ±0.43°C +0.2 / -1.5°C ±0.43°C
400 to 500°C - -0.1 / -2.9°C ±0.48°C -0.9 / -2.2°C ±0.48°C

Control Unit
Model ACS II-1000 ACS II-2000 ACS II-3000 ACS II-4000
Surface Range Rt to 300°C Rt to 500°C - Rt to 500°C
Internal Range - - Rt to 500°C Rt to 500°C
Control Principle PID auto-Tuning Control
Control Input Pt 100Ω, ø3.2 Class A
Accuracy ±(0.2% of reading)°C
Display 8mm high LED display
Reference
Sensor
Input - Pt 100Ω, ø4.8 Class A
Accuracy - ±(0.05% of reading +0.05)°C
Display - 12.5mm high LED display
Power Consumption App.160W App.700W App.1600W
Dimensions (mm) 320(W)×177(H)×350(D) 430(W)×177(H)×350(D)
Weight App.4.5kg App.5.0kg

Calibration Unit
Model ACS II-1000 ACS II-2000 ACS II-3000 ACS II-4000
Testing Surface (mm) ø78 ø100 - 150×150
Test Point - - 8 holes based on specified dimension
Temp. Distribution ±0.1°C
Heater 150W 600W 1500W
Dimensions (mm) 260(W)×130(H)×260(D) 390(W)×230(H)×360(D) 450(W)×405(H)×400(D)
Weight App.7kg App.20kg App.70kg
Overheating prevention mechanism Bimetal breaker
Heating time Rt → 150°C 20min Rt → 250°C 20min Rt → 250°C 60min
Rt → 300°C 40min Rt → 500°C 50min Rt → 500°C 150min
Cooling time 300 → 150°C 20min 500 → 250°C 30min 500 → 250°C 220min
150 → 50°C 40min 250 → 50°C 70min 250 → 100°C 430min

Data sheet

Catalogue



See also
Temperature Surface Calibrator
Anritsu Model ACSII series products are temperature calibration systems for surface temperature and internal temperature measuring devices. Temperatures can be set across a wide range, from room tempe.....
Surface Calibrator
EDL STS series of calibrators were developed to produce a traceable calibration method for surface temperature readings. In order to calibrate the STS a method was required to establish a connection .....
Dual Black Body Calibrator
EDL dual zone Black Body calibrators are designed to quickly and easily calibrate infrared thermometers with a low and high calibration point. DBB calibrators allow you to preset the required tempera.....


Back

Featured products


Barometric Pressure Transfer Standard

Vaisala Anritsu Tokico Sato Sekonic Dickson Isuzu Kobayashi
Dynasonics Ryutai RCM flowmeter Hedland Flowmeter Flomec